Automatic balancing of coaxial AC impedance bridges asks for simple, robust and efficient strategies suitable to be easily implemented on PC controllers. The paper considers the problem of balancing a bridge with n detectors to be equivalent to the problem of finding the root of a complex-valued vector function F. To avoid electrical modeling of the bridge, F is considered generally unknown except at specific points evaluated by experiment. In this framework, efficient root-finding strategies can be employed; in particular, secant method and two generalizations of it to the case n>1 are proposed. Experimental results on a test circuit are given.