Demonstrates effective benzodiazepine detection in urine, indicating a practical analytical method.
A gas diffusion microextraction (GDME) method coupled with gas chromatography–mass spectrometry (GC–MS) was developed and validated for the determination of diazepam, alprazolam, and lorazepam in urine samples. The procedure employs a headspace configuration using a low-cost cellulosic membrane as the gas-diffusion interface, allowing analyte transfer into a micro-volume methanolic acceptor phase while minimizing direct matrix contact. Key extraction parameters, including extraction temperature, time, and acceptor solvent volume, were optimized using a Box–Behnken experimental design. Under the optimized conditions (100 °C, 20 min, 200 µL methanol), the method exhibited excellent linearity in the range of 1–100 µg L−1, with coefficients of determination coefficient of determination (R2) ≥ 0.9987. The limits of detection ranged from 0.034 to 0.067 µg L−1. Intra-day and inter-day precision values were below 3.0% relative standard deviation (RSD) and the extraction recoveries were close to 100%. The analytical performance of the proposed GDME–GC–MS method was comparable to that reported for dispersive liquid–liquid microextraction (DLLME) and liquid-phase microextraction (LPME) based approaches. While the method requires elevated extraction temperature, its headspace-based configuration provides effective matrix isolation, making it a practical alternative for benzodiazepine determination in complex biological matrices such as urine.
No takes yet. Share an insight, caveat, or question.
Soltanian et al. (2026) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: