For the first time, the cryogenic radiometer of the Bureau International des Poids et Mesures (BIPM) has been taken to a national metrology institute for direct comparison with another cryogenic radiometer. This paper describes a comparison at the Physikalisch-Technische Bundesanstalt (PTB, Braunschweig, Germany) with the PTB cryogenic radiometer. The two systems have been shown to differ, in relative terms, by 2,3 × 10 -4 in the measurement of optical power at λ = 632,8 nm with a power level of about 700 μW. The standard uncertainty in the determination of this relative difference is estimated to be 5,7 × 10 -5 . As the source of this discrepancy, which is larger than the uncertainties, has not yet been found, a closer investigation of the two systems is necessary.