Paradoxical pacing spikes during a wide complex tachycardia were caused by a slow ventricular tachycardia rate below the upper tracking limit and delayed sensing due to the distance between the tachycardia exit site and the sensing lead.
Case Report (n=1)
No
Paradoxical pacing spikes during wide complex tachycardia in a CRT-D patient can result from a slow VT rate below the upper tracking limit combined with delayed sensing at the RV lead due to the tachycardia's exit site.
Clinicians should consider VT rate and sensing delays when interpreting pacing spikes; case report leaves open broader validation.
We present the case of a 68-year-old man with a cardiac resynchronization therapy-defibrillator Medtronic device implanted 3 years ago, admitted to our hospital due to a wide complex tachycardia with unexpected pacing spikes inside QRS complexes. Although that could be easily attributed to ventricular undersensing, more complicated mechanisms are involved, including the origin of the tachycardia, the parameters of the device, and the position of the device and the leads.
No takes yet. Share an insight, caveat, or question.
Leventopoulos et al. (2022) conducted a case report in Wide complex tachycardia with paradoxical pacing spikes (n=1). Biventricular defibrillator (CRT-D) was evaluated on Mechanism of paradoxical pacing spikes. Paradoxical pacing spikes during a wide complex tachycardia were caused by a slow ventricular tachycardia rate below the upper tracking limit and delayed sensing due to the distance between the tachycardia exit site and the sensing lead.