We present the realization and characterization of a new type of plastic micropump based on the magnetic actuation of a magnetic liquid. The pump consists of two serial check-valves that convert the periodic motion of a ferrofluidic plug into a pulsed quasi-continuous flow. The ferrofluid is actuated by the mechanical motion of an external NdFeB permanent magnet. The water-based ferrofluid is synthesized in-house using a coprecipitation method and has a saturation magnetization of 32 mT. The micropump consists of various layers of polymethylmethacrylate (PMMA), which are microstructured by powder blasting or by standard mechanical micromachining techniques, and are assembled in a single plastic structure using a monomer gluing solution. Two soft silicone membranes are integrated in the microfluidic structure to form two check-valves. Water has been successfully pumped at flow rates of up to 30 /spl mu/L/min and pumping is achieved at backpressures of up to 25 mbar.
No takes yet. Share an insight, caveat, or question.
Yamahata et al. (2005) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: