Why the study?
Does a combined visualization and tracking coil improve catheter tip visualization compared to a tracking coil alone in phantom and animal models?
Does a combined visualization and tracking coil improve catheter tip visualization compared to a tracking coil alone in phantom and animal models?
A novel dual-mode MR coil design improves the visualization of both the catheter tip and its distal curvature during active MR tracking.
May advance dual-mode MR coil development for tracking; leaves open clinical translation pending human studies.
In active catheter tracking, small RF coils are attached to the catheter for localization. For interactive catheter steering at vessel branchings, it is necessary to visualize not only a single point near the catheter tip but also the entire shape and orientation of the catheter's distal end. Therefore, a 35-mm-long twisted-pair RF coil was added to a 5 French intravascular catheter with a single tip-tracking coil. With the use of small nonmagnetic electronic components at the catheter tip, and a special switching circuitry outside the catheter, the coil assembly could be operated in two different modes. During MRI, the tip-tracking coil was detuned so that the MR signal was received by the visualization coil only. During tracking, detuning was switched off and the MR signal was predominantly received by the more sensitive tracking coil. The catheter was used in combination with a MR pulse sequence with automatic slice positioning so that the current imaging slice was always placed at the position of the catheter tip. Phantom and animal experiments showed that the catheter tip is better visualized with the combined approach than with a tracking coil alone.
No takes yet. Share an insight, caveat, or question.
Zuehlsdorff et al. (2004) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: