Key result
Attenuation and scatter correction improves SPECT normalcy by ~10 percentage points without compromising sensitivity.
Absolute Event Rate: 96% vs 86%
p-value: p=0.011
BACKGROUND: Soft tissue attenuation is a prominent cause of single-photon emission computed tomography (SPECT) imaging artifacts, which may result in reduced diagnostic accuracy of myocardial perfusion imaging. A method incorporating simultaneously acquired transmission data permits nonuniform attenuation correction and when incorporating scatter correction and resolution compensation may substantially reduce interpretive errors. METHODS AND RESULTS: A prospective multicenter trial was performed recruiting patients with angiographically documented coronary disease (n=96) and group of subjects with a low likelihood of disease (n=88). The uncorrected and attenuation/scatter corrected images were read independently, without knowledge of the patient's clinical data. The detection of >/=50% stenosis was similar using uncorrected perfusion data or with attenuation/scatter correction and resolution compensation (visual or visual plus quantitative analysis), 76% versus 75% versus 78%, respectively (P=NS). The normalcy rate, however, was significantly improved with this new methodology, using either the corrected images (86% vs 96%; P=0.011) or with the corrected data and quantitative analysis (86% vs 97%; P=0.007). The receiver operator characteristic curves were also found to be marginally but not significantly higher with attenuation/scatter correction than with tradition SPECT imaging. However, the ability to detect multivessel disease was reduced with attenuation/scatter correction. Regional differences were also noted, with reduced sensitivity but improved specificity for right coronary lesions using attenuation/scatter correction methodology. CONCLUSIONS: This multicenter trial demonstrates the initial clinical results of a new SPECT perfusion imaging modality incorporating attenuation and scatter correction in conjunction with 99mTc sestamibi perfusion imaging. Significant improvements in the normalcy rate were noted without a decline in overall sensitivity but with a reduction in detection of extensive coronary disease.
No takes yet. Share an insight, caveat, or question.
Hendel et al. (1999) studied Coronary artery disease (n=184). SPECT myocardial perfusion imaging with attenuation and scatter correction and resolution compensation vs. Uncorrected SPECT perfusion imaging was evaluated on Normalcy rate and detection of >=50% coronary stenosis (p=0.011). Attenuation and scatter correction in SPECT perfusion imaging improved normalcy rate (96% vs 86%; P=0.011) with similar sensitivity for >=50% stenosis (75% vs 76%; P=NS).
Synapse has enriched 5 related papers. Consider them for comparative context: