A feasible method has recently been proposed for estimating absolute blood volume (ABV) in hemodialysis (HD) patients based on intradialytic infusion of a dialysate bolus and visual assessment of the subsequent increase in relative blood volume (RBV) tracked by the dialysis machine. The aim of this study was to develop a method for more objective determination of such RBV increase to improve the accuracy of ABV estimation. We proposed a numerical algorithm consisting of interpolation and polynomial fitting of RBV signals, which we evaluated on data from 64 HD sessions in 48 patients with 240 mL of dialysate infused approximately 1 h into HD. The estimated ABV values were compared with those from a simple two-point method described previously as well as with the values calculated using the Nadler formula and the Lemmens-Bernstein-Brodsky formula. Compared to the simple method, the improved method provided higher (more plausible) estimates of ABV (median 4.79 vs 4.53 L, p < 0.001) and specific ABV (median 68.5 vs 66.4 mL/kg, p < 0.001). The improved method also provided much lower intra-patient variability of ABV estimated in different sessions of the same week (median spread 180 vs 462 mL, p < 0.001) and showed narrower limits of agreement with both Nadler and Lemmens-Bernstein-Brodsky formulae. The proposed numerical method constitutes a substantial improvement over the simple method by averaging the noise and short-term variability in RBV signals. More accurate estimates of ABV in HD patients could aid in managing their fluid status.
No takes yet. Share an insight, caveat, or question.
Pstras et al. (2023) studied this question.
Synapse has enriched 5 related papers. Consider them for comparative context: