Why the study?
Mechanical heart valves are highly thrombogenic and necessitate lifelong anticoagulation; this study evaluated hemocompatibility and thromboresistance of a novel electrically activated surface anticoagulation design.
Does an electrically activated mechanical heart valve (0.25 V or 0.5 V) improve hemocompatibility and thromboresistance compared to an uncharged valve in an in-vitro model?
Does an electrically activated mechanical heart valve (0.25 V or 0.5 V) improve hemocompatibility and thromboresistance compared to an uncharged valve in an in-vitro model?
Imposing a 0.5 V electrical potential on a mechanical heart valve reduces thrombogenic deposits in vitro without causing hemolysis or coagulation activation, establishing a foundation for future in vivo studies.
May enable anticoagulation-free MHVs; hypothesis-generating and requires in vivo trials before any clinical consideration.
Mechanical heart valves (MHVs) remain the most durable option for valve replacement, but they are highly thrombogenic and therefore necessitate life-long anticoagulation. We designed a novel MHV assembly based on "Surface Anticoagulation by Electrically induced Biomimetic Glycocalyx" (SAEBG) that imposes a weak negative potential across the valve's blood-contacting surfaces using an implantable pacemaker as an energy source. The concept is inspired by the native vascular endothelium, where the luminal glycocalyx carries a net negative surface charge that repels platelets and proteins. The present study aims to assess hemocompatibility and thromboresistance in our electrically activated MHV. Bileaflet MHVs (n = 9) were immersed in human platelet-rich plasma or whole blood, of which 5 valves were connected to a programmable pulse generator. Control valves (0 V) (n = 4) were compared to those activated at 0.25 (n = 3), and 0.5 V(n = 2). The control/activated valves were immersed in PRP/blood for 30 min under gentle agitation at 35 ± 2 °C. Hemolysis and blood cell integrity in the supernatant fluid were quantified using standard hematology analyses. Valves immersed in PRP were assessed by scanning electron microscopy (SEM). The cell/deposit free area as assessed by SEM was highest with 0.5 V (~96%), while the corresponding free areas for controls and 0.25 V were ~86% and ~58%, respectively. The platelet counts, platelet and coagulation markers were close to the measurement uncertainty ranges for all the experiments. Hemolysis was < 0.1% for all conditions, and the leucocyte and red-cell counts changed by <2%. Imposing a mild electrical potential (~0.5 V) on an MHV reproduces the anti-thrombotic behaviour of vascular endothelium. The electrical field did not injure blood cells or activate coagulation pathways, establishing a safe and effective voltage window for the SAEBG valve to enable further swine model experiments.