Answers are grounded in this paper and cite their sources.
Room temperature measurements of the ratio D T / mu between the transverse diffusion coefficient and the mobility of mass-identified ions in a neutral gas are reported for Na + ions in argon and neon. A new apparatus was used which, in contrast to previous methods, does not involve mechanical movement which alters the geometry of the drift tube assembly during a measurement scan, or variation of the experimental conditions inside it. Results with an estimated accuracy of better than 3% were obtained at E/N values ranging from 10 to 260 Td with argon and 10 to 200 Td with neon.
No takes yet. Share an insight, caveat, or question.
Hogan et al. (1990) studied this question.