OBJECTIVE: To investigate the effect of consecutive 1000-Hz tympanometry testing on admittance measures, and examine test-retest reliability. DESIGN: Repeated measures with eight trials of 1000-Hz Y(a) and B(a)/G(a) tympanometry, respectively, in two ears of each subject, followed by repeated 226-Hz tympanometry. STUDY SAMPLE: Twenty-seven normal-hearing young adults. RESULTS: For single-peak tympanograms, peak Y(tm) and G(tm) systematically increased across trials with a mean change of 8% and 15% at Trial 8, respectively, whereas B(tm) increased marginally. For notched tympanograms, peak Y(tm) and B(tm) decreased by 23% and 162% at Trial 8, and G(tm) in two cases also decreased (10% on average). Trial 2 and 3 contributed 50% to 70% of the total changes. Test-retest differences of subsequently acquired 1000- and 226-Hz tympanograms were smaller than previously reported. CONCLUSIONS: Consecutive testing significantly alters middle-ear admittance in 1000-Hz tympanometry. The outcome is contingent on tympanogram pattern and admittance component: Increase of peak Y(tm) and G(tm) in single-peak tympanograms and decrease of all measures in notched tympanograms. The present results complement previous studies on our understanding of the mechanism underlying this effect: a decrease of middle-ear stiffness. The effect of repetitive tympanometry should be accounted for in research involving sequential testing.
No takes yet. Share an insight, caveat, or question.
Carazo et al. (2014) studied this question.