Answers are grounded in this paper and cite their sources.
Stopping-power ratios for water to air, Sw,air, have been calculated for 60Co and thin-target bremsstrahlung beams. The Monte Carlo method was used to generate depth-dependent Sw,air values and depth-absorbed-dose distributions in water for monoenergetic photon beams covering the energy range 0.1-50 MeV. An averaging procedure has been developed to derive the stopping-power ratios for a photon spectrum from the monoenergetic data. Values of the Spencer-Attix ratio for delta = 10 keV, SSAw,air (delta = 10 keV), and the Bragg-Gray ratio, SBGw,air, evaluated using the new I values and two different sets of values for the density effect correction, are given at the reference depths recommended in the dosimetry protocols.
No takes yet. Share an insight, caveat, or question.
Andreo et al. (1985) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: