This is a word of caution to anyone using ionization chambers protected by thin rubber sheaths in water. Four Farmer-type ionization chambers contaminated with talcum powder were received for calibration by the Accredited Dosimetry Calibration Laboratory at the University of Texas M. D. Anderson Cancer Center. The chambers show a marked energy dependence (5% to 20%) to soft orthovoltage x rays. The response of the contaminated chambers is compared with the chambers' response before contamination and after cleaning. Techniques for identifying contaminated chambers and suggestions for cleaning them are presented.
No takes yet. Share an insight, caveat, or question.
Hanson et al. (1988) studied this question.