Does gadolinium contrast agent affect measurements of aortic flow and myocardial displacement by phase-contrast cardiovascular magnetic resonance?
Does gadolinium contrast agent affect measurements of aortic flow and myocardial displacement by phase-contrast cardiovascular magnetic resonance?
Gadolinium contrast agent does not significantly alter quantitative measurements of aortic flow or myocardial displacement by PC-CMR, though post-contrast delineation of the myocardium may be difficult due to lower image contrast.
BACKGROUND: Quantitative blood flow and aspects of regional myocardial function such as myocardial displacement and strain can be measured using phase-contrast cardiovascular magnetic resonance (PC-CMR). Since a gadolinium-based contrast agent is often used to measure myocardial infarct size, we sought to determine whether the contrast agent affects measurements of aortic flow and myocardial displacement and strain. Phase-contrast data pre and post contrast agent was acquired during free breathing using 1.5T PC-CMR. RESULTS: For aortic flow and regional myocardial function 12 and 17 patients were analysed, respectively. The difference pre and post contrast agent was 0.03±0.16 l/min for cardiac output, and 0.1±0.5 mm for myocardial displacement. Linear regression for myocardial displacement (MD) after and before contrast agent (CA) showed MDpostCA=0.95MDpreCA+0.05 (r=0.95, p<0.001). For regional myocardial function, the contrast-to-noise ratios for left ventricular myocardial wall versus left ventricular lumen were pre and post contrast agent administration 7.4±3.3 and 4.4±8.9, respectively (p<0.001). The contrast-to-noise ratios for left ventricular myocardial wall versus surrounding tissue were pre and post contrast agent administration -16.9±22 and -0.2±6.3, respectively (p<0.0001). CONCLUSIONS: Quantitative measurements of aortic flow yield equal results both in the absence and presence of gadolinium contrast agent. The total examination time may thereby be reduced when assessing both viability and quantitative flow using PC-CMR, by assessing aortic flow post contrast agent administration. Phase-contrast information for myocardial displacement is also assessable both in the absence and presence of contrast agent. However, delineation of the myocardium may be difficult or impossible post contrast agent due to the lower image contrast. Acquisition of myocardial displacement should therefore be performed pre contrast agent using current PC-CMR sequences.
No takes yet. Share an insight, caveat, or question.
Hedström et al. (2010) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: