Central venous catheters, particularly long beveled polyethylene ones, can cause lethal acute pericardial tamponade, highlighting the need for careful placement and use of pliable nonbeveled silicone elastomer catheters.
High mortality from missed CVC tamponade in case reports warrants vigilance; leaves open optimal catheter selection for prevention.
A case of acute pericardial tamponade occurred secondary to a central venous catheter; and 11 similar cases are reviewed from the literature. The tamponade in these cases is thought to be secondary to perforation of an intrapericardial structure by the central venous catheter. In 11 of the 12 patients a venous cutdown was performed using a long beveled polyethylene catheter. The correct diagnosis was made in only five instances prior to the patient's death, and in only a single case was pericardiocentesis performed early allowing for the patient's survival. Careful placement of the catheter tip outside of the heart, and the use of pliable nonbeveled silicone elastomer (Silastic) catheters are suggested as means of preventing this lethal complication of central venous catheterization.