This is the authors' abstract. We don't add key points for this paper.
An important class of analytes are volatile organic carbons (VOCs), particularly aliphatic primary alcohols. Here, we report the straightforward modification of a commercially available carbon monoxide sensor to detect a range of aliphatic primary alcohols at room temperature. The mass transport mechanisms governing the performance of the sensor were investigated using diffusion in multiple layers of the sensor to model the response to an abrupt change in analyte concentration. The sensor was shown to have a large capacitance because of the nanoparticulate nature of the platinum working electrode. It was also shown that the modified sensor had performance characteristics that were mainly determined by the condensation of the analyte during diffusion through the membrane pores. The sensor was capable of a quantitative amperometric response (sensitivity of approximately 2.2 µA/ppm), with a limit of detection (LoD) of 17 ppm methanol, 2 ppm ethanol, 3 ppm heptan-1-ol, and displayed selectivity towards different VOC functional groups (the sensor gives an amperometric response to primary alcohols within 10 s, but not to esters or carboxylic acids).
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Saunders et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: