Why the study?
Does a portable Power Doppler Imaging instrument accurately screen for high-grade carotid stenosis compared to continuous wave Doppler and routine duplex ultrasonography in high-risk patients?
Does a portable Power Doppler Imaging instrument accurately screen for high-grade carotid stenosis compared to continuous wave Doppler and routine duplex ultrasonography in high-risk patients?
Both portable Power Doppler Imaging and continuous wave Doppler are suitable as inexpensive screening strategies for identifying >75% carotid stenosis, though cw-Doppler demonstrated superior diagnostic performance.
Supports portable PDI and cw-Doppler for screening >75% carotid stenosis; hypothesis-generating and requires prospective validation before adoption.
BACKGROUND AND PURPOSE: We investigated the diagnostic performance of a brief Power Doppler Imaging (PDI) screening examination for carotid artery stenoses using a newly developed portable instrument. METHODS: A highly experienced sonographer screened in total 152 carotid arteries by either continuous wave (cw) Doppler (n= 50) or a lightweight (2.4 kg) portable duplex device (n= 102) in a prospective study of 76 high-risk patients. The screening protocols included either spectrum analysis and frequency shift measurement in both internal carotid arteries with cw-Doppler or determination of area and diameter ratios in transverse and longitudinal views of both carotid arteries in B-mode and with PDI, but without velocity measurement. Both protocols were evaluated against a complete routine duplex ultrasonography examination. RESULTS: According to the complete examination, stenoses were <50% in 73 of 102 (71.6%), 50-75% in 19 of 102 (18.6%), 75-95% in 7 of 102 (6.9%), and occluded in 3 of 102 (2.9%) arteries (PDI cohort), and <50% in 39 of 50 (78%), 50-75% in 8 of 50 (16%), 75-95% in 2 of 50 (4%), and occluded in 1 of 50 (2%) artery (cw-Doppler cohort). Mean screening time was 8.8 +/- 2.5 minute (PDI) and 9.4 +/- 2.6 minute (cw-Doppler). For stenoses >75%, A(z) values (area under the receiver operating characteristics curve) were 0.897 for area ratios, 0.843 for diameter ratios (PDI protocol) and 1.0 for the cw-Doppler protocol. CONCLUSIONS: The diagnostic performance of the cw-Doppler protocol was superior to the PDI protocol. Nevertheless, both protocols appear suitable as inexpensive screening strategies to identify subjects with >75% stenosis measured by carotid Doppler ultrasound. However, these preliminary data need further verification.
No takes yet. Share an insight, caveat, or question.
Schminke et al. (2006) studied this question.