Answers are grounded in this paper and cite their sources.
Experimental study demonstrates nanomolar protein detection in buffer and human urine using enzyme-nanoparticle sensor arrays, highlighting strong potential for diagnostic applications.
We have developed an enzyme-nanoparticle sensor array where the sensitivity is amplified through enzymatic catalysis. In this approach cationic gold nanoparticles are electrostatically bound to an enzyme (beta-galactosidase, beta-Gal), inhibiting enzyme activity. Analyte proteins release the beta-Gal, restoring activity and providing an amplified readout of the binding event. Using this strategy we have been able to identify proteins in buffer at a concentration of 1 nM, substantially lower than current strategies for array-based protein sensing. Moreover, we have obtained identical sensitivity in studies where the proteins are spiked into the complex protein matrix provided by desalted human urine ( approximately 1.5 muM total protein; spiked protein concentrations were 0.067% of the overall protein concentration), demonstrating the potential of the method for diagnostic applications.
No takes yet. Share an insight, caveat, or question.
Miranda et al. (2010) studied this question.