Why the study?
Does the Nichols Advantage automated assay provide comparable plasma aldosterone concentration measurements to a standard extraction method in patients with and without chronic renal failure?
Does the Nichols Advantage automated assay provide comparable plasma aldosterone concentration measurements to a standard extraction method in patients with and without chronic renal failure?
The automated Nichols Advantage assay for plasma aldosterone is reliable in patients with normal renal function but shows significant discrepancies compared to standard extraction methods in patients with chronic renal failure.
Discrepancies in CRF warrant assay-specific caution for PAC interpretation; cross-sectional data leaves open need for validation in renal impairment.
A variety of clinical conditions present with an abnormality in aldosterone concentration. Primary hyperaldosteronism (PHA), long considered a rare cause of hypertension (<2% of the hypertensive population), has in the past decade been suggested to be present in 3%–32% of hypertensive patients (1). Although commercially available reagent sets have made it technically easier to measure plasma aldosterone concentration (PAC) for PHA analysis, these methods are often time-consuming and limited to small throughput, with reported problems arising with their application to assay PAC from patients with chronic renal failure (CRF) (2)(3)(4). We previously validated a new automated method for measuring direct renin (5), and because there was little, if any, information on the evaluation of a similar method for aldosterone, we applied this automated method to measure PAC and compared the results with those obtained from a standard extraction method. We compared measurements of samples from hypertensive and normotensive individuals with CRF and applied samples from hypertensive or normotensive individuals without CRF as controls. We also evaluated whether different anticoagulants have any effect on PAC measured by this automated method. The study participants were of mixed ethnic origin. CRF patients were either regular hemodialysis (HD) patients with an assumed glomerular filtration rate <5 mL/min, or low-clearance (LC)-CRF patients not receiving dialysis [mean (SD) creatinine, 458 (119) μmol/L]. We obtained samples from 67 persons, of whom 37 were controls without CRF and 30 were CRF patients (19 HD and 11 LC-CRF). We obtained informed consent from all participants. We collected blood samples in both lithium-heparin (Li-Hep) and potassium EDTA tubes. For HD patients, we collected blood samples pre- and postHD. We measured PAC in all participant samples with the Nichols Advantage® (NA) fully automated assay and the standard in-house extraction method of St. Mary’s Hospital, London [described in (6) and cited in(7)]. The dynamic range for the NA assay, as given by the manufacturer, is ≤3330 pmol/L, and the limit of detection was 41 pmol/L. All results are expressed in units of pmol/L. Overall imprecision data of the NA assay, determined by measurement of 98 samples with PACs of 33.0–1671.1 pmol/L showed good precision over this interval, with lower precision at low-value samples (Fig. 1A1 ). The mean (SD) intra- and interassay CVs were 4.0 (3.2)% and 3.0 (1.8)%, respectively (n = 7). Intraassay CVs were 5.9% for a mean PAC of 247 pmol/L and 1.0% for a mean PAC of 1597 pmol/L. These values were similar to the manufacturer’s stated intraassay CVs of 5.4% at a mean PAC of 228 pmol/L and 2.9% at a mean PAC of 1990 pmol/L. We applied the NA and extraction methods to measure corresponding samples collected under different conditions of anticoagulation (EDTA and Li-Hep). With both methods, there was a strong correlation between results from EDTA vs Li-Hep plasma samples (results not shown). In patients without CRF, there was a significantly strong correlation (r2 = 0.90; P <0.0001) between PACs measured with the NA and the extraction methods. Means (SE) for the NA and extraction methods showed no significant difference at 231 (26) and 237 (21) pmol/L, respectively (Fig. 1B1 ). By contrast, there was poor agreement between the 2 methods for samples from the CRF group, especially at the high concentrations found in CRF patients (Fig. 1C1 ). The correlation (r2) between the 2 methods for PACs in LC-CRF patients (n = 11) and in CRF patients preHD (n = 18) were 0.83 (P = 0.0001) and 0.79 (P <0.0001), respectively, whereas the mean concentrations were significantly different in both cases: in the LC-CRF group the mean (SE) was 495 (92) pmol/L with the NA method and 212 (45) pmol/L with the extraction method, and in the preHD group it was 774 (181) pmol/L with the NA method and 248 (106) pmol/L with the extraction method. The correlation for PACs in post-HD samples (n = 18) was not significant (r2=0.04). In conclusion, our results show no significant difference in aldosterone measured in EDTA vs Li-Hep plasma. In patients without CRF, the aldosterone values obtained from the NA method compare well with extracted results, with only a few discrepancies. Similar to the NA method for direct renin, the automated aldosterone method has the advantage of high throughput. Therefore, the combination of these NA methods can facilitate PHA population-based studies and screening. Care must be taken with CRF patients, however, because we observed large discrepancies between PAC measured by the NA method and by extraction in patients with CRF. Precision profile and correlation of NA and extracted aldosterone results. (A), precision profile showing how imprecision with the NA method varies over a range of PACs (n = 98). Each data point represents an average percentage of CV for a range of PACs. The ranges are indicated by the dashed lines and averages are based on a minimum of 3 and a maximum of 11 sample values. (B), Bland-Altman plot showing results for all non–renal-failure samples, including 2 quality control samples (n = 39). (C), Bland-Altman plot showing results for low-clearance CRF (n = 11), preHD (n = 18), and postHD (n = 18) samples in renal failure. Two outliers (>5 SD of the residuals), a preHD and a postHD sample from the same patient, were excluded from all calculations. We thank the Supraregional Assay Service Laboratory for Aldosterone and Renin at St. Mary’s Hospital (London), in particular Mike Scanlon, as well as the Department of Chemical Pathology at St. George’s, Univerisity of London for use of their facilities and Nichols Institute Diagnostics for providing us with the reagents.
No takes yet. Share an insight, caveat, or question.
Hartman et al. (2006) studied this question.