BACKGROUND AND PURPOSE: Bone-anchored titanium implants have been used for anchorage of amputation prostheses for more than one and a half decades. Histo-logical and ultrastructural analyses were performed on a forearm amputation prosthesis after being in use for more than 11 years. MATERIAL, METHODS AND RESULTS: The implant was retrieved from the ulnar bone after a fatigue fracture of the titanium implant, and was clinically stable at the time of removal. The histological findings showed a large amount of bone within the threads and a high degree of apposition of mineralized bone to the implant surface. Ultrastructural analysis of thin samples prepared by focused ion-beam microscopy revealed an electron-dense layer at the interface and direct apposition of crystalline hydroxyapatite at the implant surface. INTERPRETATION: Our observations in this retrieval study provide a structural correlate to the functional properties and clinical results of amputation prostheses.
No takes yet. Share an insight, caveat, or question.
Palmquist et al. (2008) studied this question.