Key result
Esophageal Doppler corrected flow time tracks LV preload changes and aligns with maximal stroke volume.
Why the study?
Whether the descending aortic blood flow velocity waveform variable of flow time corrected for heart rate, measured by an esophageal Doppler transducer, can be used for noninvasive optimization of LV filling needed confirmation.
Can esophageal Doppler measurement of corrected flow time be used for noninvasive optimization of left ventricular filling compared to pulmonary arterial occlusion pressure in mechanically ventilated patients?
Observational (n=43)
Can esophageal Doppler measurement of corrected flow time be used for noninvasive optimization of left ventricular filling compared to pulmonary arterial occlusion pressure in mechanically ventilated patients?
Esophageal Doppler measurement of aortic blood flow provides a rapid, noninvasive method for optimizing left ventricular filling in mechanically ventilated patients.
Answers are grounded in this paper and cite their sources.
Corrected flow time may track LV preload changes similarly to PAOP; hypothesis-generating for noninvasive optimization in ventilated patients.
Objective To confirm whether the descending aortic blood flow velocity waveform variable of flow time corrected for heart rate, measured using an esophageal Doppler transducer, can be used for noninvasive optimization of left ventricular (LV) filling. Setting ICU and operating theater. Subjects Forty-three mechanistically ventilated patients in the ICU or undergoing cardiothoracic surgery in whom a pulmonary arterial catheter was in situ. Interventions LV preload was a) increased from hypovolemic states (pulmonary arterial occlusion pressure [PAOP] less than 8 mm Hg) by fluid challenge, b) decreased from normovolemic states (PAOP 10 to 15 mm Hg) by either iv nitrates or intravascular fluid loss, and c) decreased from heart failure or fluid overload states (PAOP greater than 20 mm Hg) by iv nitrates. No other maneuver was performed concurrently. Measurements and main results Descending aortic blood flow was measured by an esophageal Doppler transducer. Corrected flow time was calculated by dividing systolic flow time by the square root of the cycle time. PAOP and corrected flow time increased after fluid challenges in hypovolemic states, and decreased when LV preload was decreased from normovolemic states. However, when preload was decreased from overload states, PAOP always decreased, but the corrected flow time usually increased before any subsequent decrease. The greatest value of corrected flow time corresponded with the maximal stroke volume seen. Conclusions Esophageal Doppler measurement of aortic blood flow can be used for rapid, noninvasive optimization of LV filling in mechanically ventilated patients.
No takes yet. Share an insight, caveat, or question.
Singer et al. (1991) conducted an observational in Mechanically ventilated patients in the ICU or undergoing cardiothoracic surgery (n=43). Esophageal Doppler transducer measurement of corrected flow time during LV preload manipulation vs. Pulmonary arterial occlusion pressure (PAOP) was evaluated on Corrected flow time and PAOP response to LV preload alterations, and correspondence with maximal stroke volume. Esophageal Doppler corrected flow time tracked LV preload changes across PAOP thresholds (<8, 10-15, and >20 mm Hg), with maximal flow time corresponding to maximal stroke volume in 43 patients.