The demand for large and reliable data sets on isotopic composition has increased in geochemistry and environmental sciences over recent years. We present an automated ion chromatographic separation method using a robotic pipetting arm, termed ‘ ChemCobOne ’, to reduce sample separation time. Its performance was tested for lithium isotope separation in geological reference materials using a single‐step separation with HCl (0.2 mol l −1 ) and a 2 ml resin volume. This refined lithium purification method does not forfeit precision, accuracy or purity compared with manual sample processing. In addition, a δ 7 Li value for NASS‐6 of 30.99 ± 0.50‰ (2 s ) (95% CI = 0.14‰, n = 44) was determined and the first δ 7 Li values for the granite rock reference material GS‐N (−0.57 ± 0.25‰ (2 s ), 95% CI = 0.15‰, n = 15), and for the soil reference material NIST SRM 2709a (−0.37 ± 0.67‰ (2 s ), 95% CI = 0.15‰, n = 63) are proposed.
No takes yet. Share an insight, caveat, or question.
Kuessner et al. (2019) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: