PTB and DLR join their expertise and experience in optical radiometry and in THz techniques to perform what is to our knowledge the first traceable measurement of radiant power of a THz quantum cascade laser and the first absolute calibration of a THz radiation detector against a cryogenic radiometer (CR). A total standard uncertainty of 7.3% was achieved at a frequency of 2.5 THz corresponding to a wavelength of 120 µm. This uncertainty is dominated by the limited knowledge of the absorptance of the CR cavity. All other uncertainty contributions including those arising from diffraction are only 2%.