We have developed two different optical techniques for the detection of nanoscale particles. One of the methods is based on measuring the optical gradient force exerted on a nanoparticle as it passes through a confined optical field, and the other method uses a background-free interferometric scheme to detect the scattered field amplitude from a laser-irradiated particle. In both cases, the measured signal depends on the third power of the particle size (R3) as opposed to the R6dependence inherent to traditional scattering-based detection methods. The weaker size dependence in our schemes leads to a better signal-to-noise ratio (SNR) for small particles. Similar to mass spectrometry, the first detection method influences the trajectory of a particle as it passes through a tightly focused laser beam. On the other hand, the second detection method combines an interferometer with a split detector that yields no signal in the absence of a particle. For both systems, we demonstrate real-time (1 ms) detection of single nanoparticles in a microfluidic system and discuss the limits of each detection approach