To reduce the window period between HIV-1 infection and the ability to diagnose it, a fourth-generation immunoassay including the detection of HIV-1 p24 antigen has been developed. However, because the commercially available systems for this assay use special, high-cost instruments to measure, for example, chemiluminescence, it is performed only by diagnostics companies and hub hospitals. To overcome this limitation, we applied an ultrasensitive ELISA coupled with a thio-NAD cycling, which is based on a usual enzyme immunoassay without special instruments, to detect HIV-1 p24. The p24 detection limit by our ultrasensitive ELISA was 0.0065 IU/assay (i.e., ca. 10(-18) moles/assay). Because HIV-1 p24 antigen is thought to be present in the virion in much greater numbers than viral RNA copies, the value of 10(-18) moles of the p24/assay corresponds to ca. 10(3) copies of the HIV-1 RNA/assay. That is, our ultrasensitive ELISA is chasing the detection limit (10(2) copies/assay) obtained by PCR-based nucleic acid testing (NAT) with a margin of only one different order. Further, the detection limit by our ultrasensitive ELISA is less than that mandated for a CE-marked HIV antigen/antibody assay. An additional recovery test using blood supported the reliability of our ultrasensitive ELISA.