Why the study?
Does phase-shift mode radiofrequency ablation improve the creation of continuous and transmural lesions compared to bipolar and unipolar modes in a simulated 5-mm thick atrial wall?
Does phase-shift mode radiofrequency ablation improve the creation of continuous and transmural lesions compared to bipolar and unipolar modes in a simulated 5-mm thick atrial wall?
Computer simulations suggest that phase-shift mode radiofrequency ablation (90°, 135°, and 180°) creates larger, continuous, and transmural lesions compared to unipolar and bipolar modes, indicating potential utility for atrial fibrillation ablation.
Findings from finite element modeling should not alter ablation practice; leaves open need for clinical validation of PsM versus BiM/UiM.
Purpose In pulmonary veins (PVs) isolation (PVI), radiofrequency (RF) energy is often used to create a linear lesion for blocking the accessory conduction pathways around PVs. By using transient finite element analysis, this study compared the effectiveness of phase-shift mode (PsM) ablation with bipolar mode (BiM) and unipolar mode (UiM) in creating a continuous lesion and lesion depth in a 5-mm thick atrial wall. Materials and methods Computer models were developed to study the temperature distributions and lesion dimensions in atrial walls created through PsM, BiM, and UiM. Four phase-shift angles - 45°, 90°, 135°, and 180° - were considered in PsM ablation (hereafter, PsM-45°, PsM-90°, PsM-135°, and PsM-180°, respectively). Results At 60 s/30 V peak value of RF voltage, UiM and PsM-45° did not create an effective lesion, whereas BiM created a lesion of maximum depth and width approximately 1.01 and 1.62 mm, respectively. PsM-135° and PsM-180° not only created transmural lesions in 5-mm thick atrial walls but also created continuous lesions between electrodes spaced 4 mm apart; similarly, PsM-90° created a continuous lesion with a maximum depth and width of nearly 4.09 and 6.12 mm. Conclusions Compared with UiM and BiM, PsM-90°, PsM-135° and PsM-180° created continuous and larger lesions in a single ablation procedure and at 60 s/30 V peak value of RF voltage. Therefore, the proposed PsM ablation method is suitable for PVI and linear isolation at the left atrial roof for treating atrial fibrillation.
No takes yet. Share an insight, caveat, or question.
Yan et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: