Barium fluoride (BaF2) is an inorganic scintillation material used for the detection of X∕gamma radiation due to its relatively high density, equivalent atomic number, radiation hardness, and high luminescence. BaF2 has a potential capacity to be used in gamma ray timing experiments due to the prompt decay emission components. It is known that the light output from BaF2 has three decay components: two prompt of those at approximately 195 nm and 220 nm with a decay constant around 600-800 ps and a more intense, slow component at approximately 310 nm with a decay constant around 630 ns which hinders fast timing experiments. We report here the development of a fast radiation detector based on a BaF2 scintillation crystal employing a special optical filter device, a multiple reflection multi-path ultraviolet region short-wavelength pass light guides (MRMP-short pass filter) by using selective reflection technique, for which the intensity of the slow component is reduced to less than 1%. The methods used for this study provide a novel way to design radiation detector by utilizing scintillation crystal with several emission bands.
No takes yet. Share an insight, caveat, or question.
Han et al. (2013) studied this question.
Synapse has enriched 2 closely related papers on similar clinical questions. Consider them for comparative context: