Key result
The chemiluminescence-based LFA achieved 110-fold higher sensitivity, a 5.6 pg/mL cTnI detection limit, and high correlation (r=0.97) with standard analyzers.
Answers are grounded in this paper and cite their sources.
Lateral flow assays (LFAs) have become the most common biosensing platforms for point-of-care testing due to their compliance with the ASSURED (affordable, sensitive, specific, user-friendly, rapid/robust, equipment-free, and deliverable to end-users) guidelines stipulated by the World Health Organization. However, the limited analytical sensitivity and low quantitative capability of conventional LFAs, which use gold nanoparticles (AuNPs) for colorimetric labeling, have prevented high-performance testing. Here, we report the development of a highly sensitive chemiluminescence (CL)-based LFA involving AuNPs conjugated with aldehyde-activated peroxidase and antibody molecules—i.e., AuNP-(ald)HRP-Ab—as a new conjugation scheme for high-performance testing in LFAs. When paired with the CL-based signal readout modality, the AuNP-(ald)HRP-Ab conjugate resulted in 110-fold enhanced sensitivity over the colorimetric response of a typical AuNP-Ab conjugate. To evaluate the performance of the CL-based LFA, we tested it with human cardiac troponin I (cTnI; a standard cardiac biomarker used to diagnose myocardial infarction) in standard and clinical serum samples. Testing the standard samples revealed a detection limit of 5.6 pg·mL−1 and acceptably reliable precision (with a coefficient of variation of 2.3%–8.4%), according to clinical guidelines. Moreover, testing the clinical samples revealed a high correlation (r = 0.97) with standard biochemical analyzers, demonstrating the potential clinical utility of the CL-based LFA for high-performance cTnI testing.
No takes yet. Share an insight, caveat, or question.
Han et al. (2020) studied Myocardial infarction (Cardiac Troponin I detection). Chemiluminescence-based lateral flow assay using AuNP-(ald)HRP-Ab conjugate vs. Conventional colorimetric AuNP-Ab conjugate and standard biochemical analyzers was evaluated on Analytical sensitivity (detection limit) and correlation with standard biochemical analyzers for human cardiac troponin I (cTnI) (r = 0.97). The chemiluminescence-based LFA achieved 110-fold higher sensitivity, a 5.6 pg/mL cTnI detection limit, and high correlation (r=0.97) with standard analyzers.
Synapse has enriched 10 related papers. Consider them for comparative context: