Key result
Oversized ASD occluder shows incomplete endothelialization and thrombus 2 years post-implant.
Why the study?
Incomplete endothelialization of atrial septal occluders and its imaging characteristics remain inadequately described in clinical practice.
Case Report (n=1)
No
Answers are grounded in this paper and cite their sources.
Multimodality imaging demonstrates that relative device oversizing and partial embedding of an atrial septal occluder can constrain expansion and hinder endothelialization, leading to persistent intra-device flow and thrombus.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Li et al. (2026) conducted a case report in Atrial septal defect (n=1). Atrial septal occluder (Cardi-O-Fix ASD Occluder) was evaluated on Device endothelialization and intra-device flow. Multimodality imaging revealed incomplete endothelialization, persistent intra-device flow, and thrombus up to 2 years after atrial septal defect closure with an oversized occluder.