Key result
An artificial neural network classifier based on noninvasive impeller speed signals detected the suction state in implantable rotary blood pumps with 98.54% sensitivity and 99.26% specificity.
Why the study?
Does an artificial neural network classifier accurately detect suction states in heart failure patients with implantable rotary blood pumps?
Does an artificial neural network classifier accurately detect suction states in heart failure patients with implantable rotary blood pumps?
Effect estimate: Sensitivity 98.54%, Specificity 99.26%
An artificial neural network can accurately detect ventricular suction in implantable rotary blood pumps using noninvasive impeller speed signals, potentially improving control strategies and clinical care.
May facilitate real-time suction avoidance in LVAD patients; leaves open prospective outcome trials.
Granting those heart failure patients who are recipients of an implantable rotary blood pump (iRBP) greater functionality in daily activities is a key long-term strategy currently being pursued by many research groups. A reliable technique for noninvasive detection of the various pumping states, most notably that of ventricular collapse or suction, is an essential component of this strategy. Presented in this study is such a technique, whereby various indicators are derived from the noninvasive pump feedback signals, and a suitable computational methodology developed to classify the pumping states of interest. Clinical telemetry data from ten implant recipients was categorized (with the aid of trans-oesophageal echocardiography) into the normal and suction states. These data are used to develop a pumping state classifier based on an artificial neural network (ANN). Nine indices, derived from the noninvasive impeller speed signal, form the inputs to this ANN classifier. During validation, the resulting ANN classifier achieved a maximum sensitivity of 98.54% (609/618 samples of 5 s in length) and specificity of 99.26% (12,123/12,213 samples) for correct detection of the suction state. The ability to detect the suction state with such a high degree of accuracy provides a critical parameter both for control strategy development, and for clinical care of the implant recipient.
No takes yet. Share an insight, caveat, or question.
Karantonis et al. (2008) studied Heart failure with an implantable rotary blood pump (n=10). Artificial neural network (ANN) classifier vs. Trans-oesophageal echocardiography (reference standard) was evaluated on Correct detection of the suction state (Sensitivity 98.54%, Specificity 99.26%). An artificial neural network classifier based on noninvasive impeller speed signals detected the suction state in implantable rotary blood pumps with 98.54% sensitivity and 99.26% specificity.