Key result
Supraclavicular nonimaging Doppler can falsely elevate aortic gradients due to brachiocephalic stenosis.
Case Report (n=1)
No
Clinicians using a nonimaging continuous-wave Doppler probe from the right supraclavicular window after TAVR should be aware that peripheral vascular stenoses in the brachiocephalic or subclavian arteries can produce falsely elevated pressure gradients mimicking prosthetic valve dysfunction or thrombosis.
Unexpectedly high post-TAVR gradients from right supraclavicular NI-CWD may reflect brachiocephalic stenosis rather than valve dysfunction; leaves open the frequency of this artifact.
•Elevated gradients after TAVR were seen using a nonimaging CWD probe.•Normal valve function was seen on transesophageal echocardiography.•A retrospective review of CT imaging showed stenosis of the brachiocephalic artery.•Capturing an elevated gradient from another lesion is a pitfall of nonimaging probes.
No takes yet. Share an insight, caveat, or question.
Lander et al. (2020) conducted a case report in Severe aortic stenosis post-transcatheter aortic valve replacement (TAVR) with brachiocephalic and right subclavian artery stenosis (n=1). Nonimaging continuous-wave Doppler (NI-CWD) probe from the right supraclavicular fossa vs. Transesophageal echocardiography (TEE) and apical transthoracic Doppler was evaluated on Aortic valve peak and mean pressure gradients. Nonimaging continuous-wave Doppler from the right supraclavicular window recorded falsely elevated peak/mean gradients (91/58 mm Hg vs 15/9 mm Hg on TEE) due to brachiocephalic stenosis.