Answers are grounded in this paper and cite their sources.
The use of a transthoracic bioimpedance monitor to determine cardiac output was evaluated in critically ill children. The children ranged in age from 10 months to 8 yr and their height and weight ranged from the third to the 97th percentile. Each child had a thermodilution catheter in place to monitor cardiac output. The bioimpedance monitor used in this study, the NCCOM-3, required the input of a constant (L), which was obtained for each individual patient by adjusting the L setting until cardiac output measured by bioimpedance (COBI) was within 10% of cardiac output measured by thermodilution (COTD). This method of determining L was superior to using either measured thoracic length or the manufacturer's guidelines to obtain L and resulted in an excellent correlation between COTD and COBI (r = .94; p less than .05; n = 59). In children less than 125 cm in height, measured thoracic length alone was inadequate to use for L but provided a good approximation of L when multiplied by 1.25. This study suggests that the use of transthoracic bioimpedance to determine cardiac output compares favorably with thermodilution techniques and it is noninvasive.
No takes yet. Share an insight, caveat, or question.
Introna et al. (1988) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: