The developed cardiorespiratory mechanical simulator reliably reproduces cardiac and respiratory movements for in vitro testing of pacemaker impedance minute ventilation sensors.
New in vitro simulator enables isolated iMV sensor testing; leaves open clinical translation pending in vivo validation.
We developed a cardiorespiratory mechanical simulator (CRMS), a system able to reproduce both the cardiac and respiratory movements, intended to be used for in vitro testing of impedance minute ventilation (iMV) sensors in cardiac pacemakers. The simulator consists of two actuators anchored to a human thorax model and a software interface to control the actuators and to acquire/process impedance signals. The actuators can be driven separately or simultaneously to reproduce the cardiac longitudinal shortening at a programmable heart rate and the diaphragm displacement at a programmable respiratory rate (RR). A standard bipolar pacing lead moving with the actuators and a pacemaker case fixed to the thorax model have been used to measure impedance (Z) variations during the simulated cardiorespiratory movements. The software is able to discriminate the low-frequency component because of respiration (Z(R)) from the high-frequency ripple because of cardiac effect (Z(C)). Impedance minute ventilation is continuously calculated from Z(R) and RR. From preliminary tests, the CRMS proved to be a reliable simulator for in vitro evaluation of iMV sensors. Respiration impedance recordings collected during cardiorespiratory movements reproduced by the CRMS were comparable in morphology and amplitude with in vivo assessments of transthoracic impedance variations.
No takes yet. Share an insight, caveat, or question.
Marcelli et al. (2015) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: