The application of bismuth film electrodes to the determination of cobalt by constant current adsorptive stripping potentiometry with exploitation of a catalytic effect is presented. The addition of NaNO2 to the solution containing ammonia buffer and dimethylglyoxime results in a 25‐fold enhancement of the adsorptive stripping potentiometric cobalt signal. Several key parameters of the potentiometric stripping mode were optimized, including the composition of the supporting electrolyte, the stripping current, the accumulation potential, and the accumulation time. The optimized procedure yields favorable and highly stable stripping responses with good precision (RSD=1.4% for a Co concentration of 2 µg L−1), low detection limit (0.07 µg L−1), and good linearity (up to 10 µg L−1, R2=0.998) with a deposition time of 60 s and a stripping current of 10 µA. The method enables the determination of Co in the presence of high excesses of Ni or Zn.