Answers are grounded in this paper and cite their sources.
Fluorescence detection of fluorogen-labeled neurotransmitters is demonstrated using 100 fs pulses from a titanium-sapphire mode-locked laser to achieve molecular excitation by simultaneous absorption of two and three photons of near-IR radiation. Two-photon excitation spectra are determined for the naphthalene-2,3 dicarboxaldehyde derivative of glycine and the fluorescamine derivative of leucine enkephalin, with the peak excitation cross section (o2) approximately equal to 1 x 10(-50) cm4 s/photon for both species. Three-photon-excitation fluorescence is demonstrated for o-phthaldialdehyde-labeled glutamate using excitation wavelengths between 965 and 1012 nm. The three-photon excitation cross section (o3) remains nearly constant in this wavelength range, with an absolute value of approximately 10(-84)-10(-85) cm6 s2/photon 2. Rapid cycling of analytes through the fluorescent excited state and detection that is free from background caused by Rayleigh and Raman scatter combine to make multiphoton-excited fluorescence a highly sensitive approach for detecting trace amounts of neurotransmitters. Measurements of two-photon-excited fluorescence of fluorescamine-labeled bradykinin and analysis of multiphoton-excited background reveal the potential of this method to detect fewer than 1000 neurotransmitter molecules.
No takes yet. Share an insight, caveat, or question.
Shear et al. (1996) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: