Answers are grounded in this paper and cite their sources.
Metal implants produce susceptibility artefacts in magnetic resonance imaging. We have explored the effects of scaphoid screw characteristics and orientation on MR susceptibility artefact. Titanium alloy, smallness and longitudinal alignment with the z-axis of the main magnetic field reduce the size of the susceptibility artefact.
No takes yet. Share an insight, caveat, or question.
Ganapathi et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: