With the objective of improving MR endoluminal imaging of the colonic wall, electromagnetic simulations of different configurations of single-layer and double-layer, and double-turn endoluminal coil geometries were run. Indeed, during colon navigation, variations in coil orientation with respect to B0 are bound to occur, leading to impaired image acquisition due to a loss of signal uniformity. In this work, three typical coil orientations encountered during navigation were chosen and the resulting signal uniformity of the different geometries was investigated through the simulated <mfenced open="(" close=")" separators="|">B1x,y/I<msqrt>Rt</msqrt></mfenced> values. Sampling this quantity over a circle of radius r enabled us to calculate the coefficient of variation (= standard deviation/mean) for this given distance. This procedure was repeated for r∈<mfenced open="[" close="]" separators="|">5;15</mfenced> mm, which represents the region of interest in the colon. Our results show that single-loop and double-layer geometries could provide complementary solutions for improved signal uniformity. Finally, using four microelectromechanical system switches, we proposed the design of a reconfigurable endoluminal coil able to switch between those two geometries while also ensuring the active decoupling of the endoluminal coil during the RF transmission of an MR experiment.