Answers are grounded in this paper and cite their sources.
AIM: Exact knowledge of dosimetric parameters is an essential pre-requisite of an effective treatment in radiotherapy. In order to fulfill this consideration, different techniques have been used, one of which is Monte Carlo simulation. MATERIALS AND METHODS: This study used the MCNP-4C to simulate electron beams from Neptun 10 PC medical linear accelerator. Output factors for 6, 8 and 10 MeV electrons applied to eleven different conventional fields were both measured and calculated. RESULTS: The measurements were carried out by a Wellhofler-Scanditronix dose scanning system. Our findings revealed that output factors acquired by MCNP-4C simulation and the corresponding values obtained by direct measurements are in a very good agreement. CONCLUSION: In general, very good consistency of simulated and measured results is a good proof that the goal of this work has been accomplished.
No takes yet. Share an insight, caveat, or question.
Toossi et al. (2012) studied this question.
Synapse has enriched 2 related papers. Consider them for comparative context: