Key result
Rapid-deployment SAVR yields ~98% freedom from structural valve deterioration at 7 years.
Why the study?
Rapid-deployment valves reduce operative time for surgical aortic valve replacement, but long-term core laboratory-adjudicated data are scarce.
What are the 7-year clinical durability and core laboratory-adjudicated echocardiographic hemodynamic outcomes of surgical aortic valve replacement using the Edwards INTUITY Elite rapid-deployment valve in patients with severe aortic stenosis?
Cohort (n=885)
nonrandomized
Yes
What are the 7-year clinical durability and core laboratory-adjudicated echocardiographic hemodynamic outcomes of surgical aortic valve replacement using the Edwards INTUITY Elite rapid-deployment valve in patients with severe aortic stenosis?
Surgical AVR using the INTUITY rapid-deployment bioprosthetic valve demonstrates durable 7-year valve function (97.5% freedom from structural valve deterioration), stable hemodynamics, and overall survival comparable to an age- and sex-matched US population, though with a 21.8% 7-year cumulative pacemaker rate in isolated AVR.
Supports 7-year durability of INTUITY rapid-deployment AVR; extends single-arm cohort data but leaves randomized comparisons open.
Background The use of rapid-deployment valves (RDVs) has been shown to reduce the operative time for surgical aortic valve replacement (AVR). Long-term core laboratory–adjudicated data are scarce, however. Here we report final 7-year data on RDV use. Methods TRANSFORM was a prospective, nonrandomized, multicenter, single-arm trial implanting a stented bovine pericardial valve with an incorporated balloon-expandable sealing frame. A prior published 1-year analysis included 839 patients from 29 centers. An additional 46 patients were enrolled and implanted, for a total of 885 patients. Annual clinical and core laboratory–adjudicated echocardiographic outcomes were collected through 8 years. Primary endpoints were structural valve deterioration (SVD), all-cause reintervention, all-cause valve explantation, and all-cause mortality. Secondary endpoints included hemodynamic performance assessed by echocardiography. The mean duration of follow-up was 5.0 ± 2.0 years. Results The mean patient age was 73.3 ± 8.2 years. Isolated AVR was performed in 62.1% of the patients, and AVR with concomitant procedures was performed in 37.9%. Freedom from all-cause mortality at 7 years was 76.0% for isolated AVR and 68.2% for concomitant AVR. Freedom from SVD, all-cause reintervention, and valve explantation at 7 years was 97.5%, 95.7%, and 97.8%, respectively. The mean gradient and effective orifice area at 7 years were 11.1 ± 5.3 mm Hg and 1.6 ± 0.3 cm 2 , respectively. Paravalvular leak at 7 years was none/trace in 88.6% and mild in 11.4%. In patients undergoing isolated AVR, the cumulative probability of pacemaker implantation was 13.9% at 30 days, 15.5% at 1 year, and 21.8% at 7 years. Conclusions AVR for aortic stenosis using an RDV is associated with low rates of late adverse events. This surgical pericardial tissue platform provides excellent and stable hemodynamic performance through 7 years.
No takes yet. Share an insight, caveat, or question.
Malaisrie et al. (2023) conducted a cohort in Severe aortic stenosis or stenosis-insufficiency necessitating primary surgical aortic valve replacement (n=885). Surgical aortic valve replacement using the Edwards INTUITY Elite rapid-deployment valve system vs. Single-arm (survival compared to an age- and sex-matched US general population cohort) was evaluated on Freedom from structural valve deterioration (SVD), all-cause reintervention, all-cause valve explantation, and all-cause mortality at 7 years. Surgical AVR with the INTUITY rapid-deployment valve achieved 7-year freedom from structural valve deterioration of 97.5%, reintervention of 95.7%, and overall survival of 73.2%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: