BACKGROUND: Frameless stereotactic surgery utilizing fiducial-based (FB) registration is an established tool in the armamentarium of deep brain stimulation (DBS) surgeons. Fiducial-less (FL) registration via intraoperative CT, such as the O-arm, has been routinely used in spine surgery, but its accuracy for DBS surgery has not been studied in a clinical setting. OBJECTIVE: We undertook a study to analyze the accuracy of the FL technique in DBS surgery and compare it to the FB method. METHODS: In this prospective cohort study, 97 patients underwent DBS surgery using the NexFrame and the O-arm registration stereotactic system. Patients underwent FB (n = 50) registration from 2015 to 2016 and FL (n = 47) O-arm registration from 2016 to 2017. RESULTS: The radial errors (RE) and vector/euclidean errors of FB and FL registration were not significantly different. There was no difference in additional passes between methods, but there was an increase in the number of RE ≥2.5 mm in the FL method. CONCLUSION: Although there was no statistically significant difference in RE or the need for additional passes, the increased number of errors ≥2.5 mm with the FL method (17 vs. 4% in FB) indicates the need for further study. We concluded that O-arm images of the implants should be utilized to assess and correct for this error.