We report here an efficient photocatalytic amperometric sensor for the determination of dissolved oxygen (DO) in phosphate buffer solution using a disposable copper-plated screen-printed carbon electrode (CuSPE). The photoelectrochemical activity toward DO of the CuSPE was related to the formation of a p-type semiconductor Cu(I)2O. The solution pH and biased potential (E(bias)) were systematically optimized as pH 8 PBS and -0.7 V vs Ag/AgCl, respectively. Under optimized conditions, the calibration plot was linear in the range of 1-8 ppm with sensitivity and regression coefficient of 23.51 (microA cm2)(-1) ppm(-1) and 0.9982, respectively. The reproducibility of the system was good with seven successive measurements of DO yielding a RSD value of 1.87%. Real sample assays for groundwater and tap water were also consistent with those measured by a commercial DO meter. The principle used in DO measurement has an opportunity to extend into various research fields.