Why the study?
How accurate and precise are pulse oximeters under various clinical and physiological conditions?
How accurate and precise are pulse oximeters under various clinical and physiological conditions?
Pulse oximeters are generally accurate at normal oxygen saturations but exhibit significant errors at saturations below 80% and during exercise with ear probes.
Answers are grounded in this paper and cite their sources.
Clinicians should interpret SpO2 <80% with caution; extends evidence on pulse oximeter inaccuracy in severe hypoxia.
The published studies of pulse oximeter performance under conditions of normal, high and low saturation, exercise, poor signal quality and cardiac arrhythmia are reviewed. Most pulse oximeters have an absolute mean error of less than 2% at normal saturation and perfusion; two-thirds have a standard deviation (SD) of less than 2%, and the remainder an SD of less than 3%. Some pulse oximeters tend to read 100% with fractional saturations of 97-98%. Pulse oximeters may be suitable hyperoxic alarms for neonates if the alarm limit chosen is directly validated for each device. Pulse oximeters are poorly calibrated at low saturations and are generally less accurate and less precise than at normal saturations; nearly 30% of 244 values reviewed were in error by more than 5% at saturations of less than 80%. Ear, nose and forehead probes respond more rapidly to rapid desaturation than finger probes, but are generally less accurate and less precise. Ear oximetry may be inaccurate during exercise. Low signal quality can result in failure to present a saturation reading, but data given with low signal quality warning messages are generally no less accurate than those without. Cardiac arrhythmias do not decrease accuracy of pulse oximeters so long as saturation readings are steady.
No takes yet. Share an insight, caveat, or question.
Webb et al. (1991) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: