Key result
Laser Doppler Vibrometry detected stenosis-induced flow instabilities in an in vitro carotid model with sensitivity comparable to intra-arterial pressure measurements, particularly in the 50-150 Hz range.
Why the study?
Early detection of asymptomatic carotid stenosis may help identify individuals at risk of stroke, prompting exploration of non-contact laser Doppler vibrometry (LDV) to detect stenosis from neck skin vibrations.
Does Laser Doppler Vibrometry detect stenosis-induced disturbed flow in an in vitro carotid bifurcation model?
Does Laser Doppler Vibrometry detect stenosis-induced disturbed flow in an in vitro carotid bifurcation model?
Laser Doppler Vibrometry can detect high-frequency fluctuations caused by stenosis-induced disturbed flow in an in vitro model, suggesting potential for non-contact cardiovascular screening.
LDV may enable non-contact carotid stenosis detection; leaves open clinical utility pending human validation.
Early detection of asymptomatic carotid stenosis may help identifying individuals at risk of stroke. We explore a new method based on laser Doppler vibrometry (LDV) which could allow the non-contact detection of stenosis from neck skin vibrations due to stenosis-induced flow disturbances. Experimental fluid dynamical tests were performed with water on a severely stenosed patient-specific carotid bifurcation model. Measurements were taken under various physiological flow regimes both in a compliant and stiff-walled version of the model, at 1 to 4 diameters downstream from the stenosis. An inter-arterial pressure catheter was positioned as reference. Increasing flow led to corresponding increase in power spectral density (PSD) of pressure and LDV recordings in the 0-500 Hz range. The stiff model lead to higher PSD. PSD of the LDV signal was less dependent on the downstream measurement location than pressure. The strength of the association between PSD and flow level, model material and measuring location was highest in the 0-50 Hz range, however useful information was found up to 200 Hz. This proof-of-concept suggests that LDV has the potential to detect stenosis-induced disturbed flow. Further computational and clinical validation studies are ongoing to assess the sensitivity and specificity of the technique for clinical screening.
No takes yet. Share an insight, caveat, or question.
Mancini et al. (2019) studied Carotid stenosis. Laser Doppler Vibrometry vs. Intra-arterial pressure catheter was evaluated on Power spectral density of skin vibrations in the 0-500 Hz range. Laser Doppler Vibrometry detected stenosis-induced flow instabilities in an in vitro carotid model with sensitivity comparable to intra-arterial pressure measurements, particularly in the 50-150 Hz range.