INTRODUCTION: Hand-held dynamometry (HHD) is commonly used to measure ankle plantarflexion strength but has variable reliability measuring higher forces. Fixed HHD is suggested to improve reliability. We, therefore, compared the reliability, consistency, and accuracy of measuring plantarflexion strength. METHODS: ), SEM, minimal detectable change, and Spearman correlation coefficients were calculated to assess inter-trial repeatability, consistency, agreement, and accuracy. RESULTS: 0.96 to 0.98) and highly correlated (Spearman rho = 0.815; P < 0.01). Fixed HHD produced significantly higher force outputs. HHD alone provided more consistent force values. CONCLUSIONS: Both methods of measuring ankle plantarflexion force were reliable. Force measured with fixed HHD will likely be more accurate for adults and individuals with greater strength, while HHD alone will be more consistent for individuals with lower strength. Muscle Nerve 56: 896-900, 2017.