Recent achievements in active capsule endoscopy have allowed controlled inspection of the bowel by magnetic guidance. Capsule localization represents an important enabling for such kinds of platforms. In this paper, the authors present a localization, applied as first step in time-discrete capsule position detection, that is useful for establishing a magnetic link at the beginning of an endoscopic procedure or for re-linking the capsule in the case of loss due to locomotion. The novelty of this approach consists in using magnetic sensors on board the capsule whose output is combined with pre-calculated magnetic analytical model solutions. A magnetic field triangulation algorithm is used for obtaining the position of the capsule inside the gastrointestinal tract. Experimental validation has demonstrated that the proposed procedure is stable, accurate and has a wide localization range in a volume of about 18 × 103 cm3. Position errors of 14 mm along the X direction, 11 mm the Y direction and 19 mm along the Z direction were obtained in less than 27 s of time. The proposed approach, being compatible with magnetic fields used for locomotion, can be easily extended to other platforms for active capsule endoscopy.