Epilepsy affects more than 1% of the world's population. Responsive neurostimulation is emerging as an alternative therapy for the 30% of the epileptic patient population that does not benefit from pharmacological treatment. Efficient seizure detection algorithms will enable closed-loop epilepsy prostheses by stimulating the epileptogenic focus within an early onset window. Critically, this is expected to reduce neuronal desensitization over time and lead to longer-term device efficacy. This work presents a novel event-based seizure detection algorithm along with a low-power digital circuit implementation. Hippocampal depth-electrode recordings from six kainate-treated rats are used to validate the algorithm and hardware performance in this preliminary study. The design process illustrates crucial trade-offs in translating mathematical models into hardware implementations and validates statistical optimizations made with empirical data analyses on results obtained using a real-time functioning hardware prototype. Using quantitatively predicted thresholds from the depth-electrode recordings, the auto-updating algorithm performs with an average sensitivity and selectivity of 95.3 +/- 0.02% and 88.9 +/- 0.01% (mean +/- SE(alpha = 0.05)), respectively, on untrained data with a detection delay of 8.5 s [5.97, 11.04] from electrographic onset. The hardware implementation is shown feasible using CMOS circuits consuming under 350 nW of power from a 250 mV supply voltage from simulations on the MIT 180 nm SOI process.
No takes yet. Share an insight, caveat, or question.
Raghunathan et al. (2009) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: