Key result
In vitro testing of On-X prosthetic mitral valves demonstrated an effective orifice area of 2.1 cm2, a performance index of 0.56 at 3.5 L/min cardiac output, and regurgitant volumes below 6 ml.
Population
In vitro computer-controlled mock loop system testing two samples of the On-X Mitral 27/29 and one sample of…
Design
Preclinical
Authors
Loading...
In vitro hydrodynamics alone do not guide valve selection; leaves open in vivo validation before clinical consideration.
In vitro mock loop testing demonstrates that On-X prosthetic mitral valves have excellent hydrodynamic characteristics, including an effective orifice area of 2.1 cm2 and low regurgitant volumes.
Verdonck et al. (2002) studied Prosthetic mitral valve testing (n=3). On-X prosthetic mitral valve was evaluated on Hydrodynamic characteristics (effective orifice area, performance index, regurgitant volume). In vitro testing of On-X prosthetic mitral valves demonstrated an effective orifice area of 2.1 cm2, a performance index of 0.56 at 3.5 L/min cardiac output, and regurgitant volumes below 6 ml.