Key result
Pacemaker firmware upgrades carry ~0.2% risk of back-up mode pacing, requiring shared decision-making.
The decision to proceed with pacemaker firmware upgrades for cybersecurity requires shared decision-making, weighing the theoretical risk of an exploit against the known small risk of device malfunction during the update process.
In August 2017, the first major recall for cybersecurity vulnerabilities in pacemakers capable of remote connectivity was released that impacted 465,000 United States patients.1,2 The FDA approved a firmware update designed by the manufacturer of the devices as a remediation (Abbott , formally St. Jude Medical). The recall was in response to the public disclosure of vulnerability by an investment firm and produced in a lab environment that could allow an unauthorized party in close proximity to a patient to impact the performance of the device or modify device settings via radio frequency communication.3 While an exploit has not occurred in a patient and requires a high degree of resources and skill to execute, if accomplished, it could pose a significant risk to device safety and essential performance and cause patient harm. The FDA defines this as an uncontrolled vulnerability.2 The recall recommendations were coordinated between the FDA, the Industrial Control Systems Cyber Emergency Response Team (ICS-CERT), a division of Homeland Security that responds to and coordinates disclosure of critical infrastructure cybersecurity vulnerabilities and Abbott.1 All parties urged caution and shared decision making between patient and clinician as to whether to have the device firmware update, a process that requires a clinic visit to implement with a device programmer. The manufacturer bench tested the firmware update but the only prior experience with an implanted device firmware updates was a 2012 ICD firmware update that demonstrated a 0.197% risk of device back-up mode pacing after the upgrade was performed.
No takes yet. Share an insight, caveat, or question.
Saxon et al. (2018) conducted a review in Cybersecurity vulnerabilities in implanted pacemakers. Firmware upgrade was evaluated. Firmware upgrades for pacemakers impacted 465,000 US patients, requiring shared decision-making to balance theoretical hacking risks against a historical 0.197% risk of device back-up mode pacing.
Synapse has enriched 5 related papers. Consider them for comparative context: