Heparinized dog blood was exposed to shear and foreign surfaces in conicylindrical test cells. The cells were injection molded from polycarbonate and were filled using a technique that avoided contact of the blood with air. Particulate-matter formation was measured and was found to be dominated by the surface-to-blood-volume ratio and to be independent of shear rate. Hemolysis was also measured and was found to vary linearly with shear rate and to increase with increasing surface-to-blood volume ratio. Thus, at low shear rates and high specific surface conditions, the degree of hemolysis was found to be minimal while particulate-matter formation was high. The results suggest that the safety of extracoporeal perfusion procedures cannot be inferred from hemolysis measurements alone. In one series of tests, a gas-blood interface was generated at a rate equivalent to the rate of surface renewal in conventional disc oxygenators. The gas-blood interface failed to contribute significantly to the damage indices, which suggests that the apparent superiority of membrane oxygenators may be a result of factors other than the absence of a blood-gas interface.
No takes yet. Share an insight, caveat, or question.
Solen et al. (1978) studied this question.
Synapse has enriched 3 closely related papers on similar clinical questions. Consider them for comparative context: