Answers are grounded in this paper and cite their sources.
The effects of analyte phase on the calibration response for laser-induced breakdown spectroscopy is investigated for a range of carbon species. Significant differences in the atomic emission signal from carbon were observed when comparing calibration streams of gas-phase and submicrometer-sized solid-phase carbon species. The resulting calibration curve slopes varied by a factor of 8 over a comparable range of atomic carbon concentrations for five different analyte sources, while the plasma electron density and temperature remained essentially constant. The current findings challenge a widely held assumption that complete dissociation of constituent species within a highly energetic laser-induced plasma results in independence of the analyte atomic emission signal on the analyte source. A physical model of the plasma-analyte interaction is proposed that provides a framework to account for the observed dependence on the physical state of the analyte.
No takes yet. Share an insight, caveat, or question.
Hohreiter et al. (2005) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: